Automatic Intrinsic Calibration of Double-sided Silicon Strip Detectors

Author

  • M. Reese
  • J. Gerl
  • Pavel Golubev
  • N. Pietralla

Summary, in English

Abstract A reliable and simple-to-use algorithm was developed for the energy-calibration of double-sided silicon strip detectors (DSSSDs). It works by enforcing mutual consistency of p-side and n-side information for every detected event. The procedure does not rely on a dedicated data set for calibration and is robust enough to work fully automated without human supervision. The method was developed and applied to data from a DSSSD of the Lund-York-Cologne CAlorimeter (LYCCA) for the HISPEC experiment at FAIR. It has been tested on ions in the A â 90 mass range at energies of E kin â 300 MeV / u .

Department/s

  • Nuclear physics

Publishing year

2015

Language

English

Pages

63-68

Publication/Series

Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment

Volume

779

Document type

Journal article

Publisher

Elsevier

Topic

  • Accelerator Physics and Instrumentation

Keywords

  • Silicon strip detector
  • Calibration procedure
  • Energy loss measurement