Mikael Elfman
Researcher

The Lund Nuclear Microprobe sub-micron set-up. Part II: Beam line, focusing system and scanning
Author
Summary, in English
A new beam line for generating a sub-micron beam spot for high-resolution nuclear microprobe applications has been constructed at the Lund Nuclear Microprobe facility. The design and construction of the beam line, the quality test of the four high precision Oxford OM-52 Quadrupole magnets by the grid shadow method and the beam scanning system are presented in this paper. The beam line is designed for a two-stage focusing system using an independent doublet at each stage. The first stage focuses the beam in an intermediate chamber, which has a scanning system and a high resolution viewing system. The beam spot in the intermediate chamber is then used as the object for the second stage. Slit construction, vacuum system and beam control is discussed. Data illustrating the quadrupole quality and scanning capability are shown.
Department/s
- Nuclear physics
Publishing year
2005
Language
English
Pages
7-13
Publication/Series
Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms
Volume
231
Issue
1-4
Document type
Journal article
Publisher
Elsevier
Topic
- Subatomic Physics
Keywords
- sub-micron beam
- quadrupole doublets
- beam scanning
- Nuclear Microprobe
Status
Published
Research group
- Nuclear Microprobe
ISBN/ISSN/Other
- ISSN: 0168-583X